Operating Instructions
Incl. Declaration of Conformity
Compact Pirani Capacitance
Gauge
PCR 260
BG 805 180 BE / C (2006-05)
Product Identification
In all communications with Pfeiffer Vacuum, please specify
the information given on the product nameplate. For conven-
ient reference copy that information into the diagram below.
Typ:
No:
F-No:
V W
Pfeiffer Vacuum, D-35614 Asslar
Validity
This document applies to products with the following part
numbers:
PT R26 850
(DN 16 ISO-KF)
PT R26 851 (DN 16 CF-F)
The part number (No) can be taken from the product
nameplate.
If not indicated otherwise in the legends, the illustrations in
this document correspond to DN 16 ISO-KF vacuum con-
nection. They apply to other vacuum connections by analogy.
We reserve the right to make technical changes without prior
notice.
All dimensions are indicated in mm.
Intended Use
The Compact Pirani Capacitance Gauge PCR 260 has been
designed for vacuum measurement of gases in the pressure
range of 5×10
-4
… 1500 mbar.
It must not be used for measuring flammable or combustible
gases which react in air.
The gauge can be operated in connection with a Pfeiffer
Vacuum controller or with another evaluation unit.
Functional Principle
The PCR gauge is a combination gauge consisting of a Pirani
sensor and a capacitive diaphragm sensor. Both sensors are
constantly active.
At low pressures, only the signal of the Pirani sensor is used
for pressure measurement; at high pressures, only the signal
of the capacitive diaphragm sensor. To determine the output
signal in the intermediate range, both signals are used
proportional to the pressure.
Safety
Symbols Used
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment and envi-
ronmental damage.
Caution
Information on correct handling or use. Disregard can lead
to malfunctions or minor equipment damage.
Personnel Qualifications
Skilled personnel
All work described in this document may only be carried out
by persons who have suitable technical training and the
necessary experience or who have been instructed by the
end-user of the product.
General Safety Instructions
Adhere to the applicable regulations and take the neces-
sary precautions for the process media used.
Consider possible reactions between the product materials
and the process media.
Consider possible reactions of the process media due to
the heat generated by the product.
Adhere to the applicable regulations (e.g. explosion) and
take the necessary precautions for all work you are going
to do and consider the safety instructions in this
document.
Before beginning to work, find out whether any vacuum
components are contaminated. Adhere to the relevant
regulations and take the necessary precautions when
handling contaminated parts.
Communicate the safety instructions to all other users.
Liability and Warranty
Pfeiffer Vacuum assumes no liability and the warranty
becomes null and void if the end-user or third parties
disregard the information in this document
use the product in a non-conforming manner
make any kind of changes (modifications, alterations etc.)
to the product
use the product with accessories not listed in the product
documentation.
The end-user assumes the responsibility in conjunction with
the process media used.
Gauge failures due to contamination, as well as expendable
parts (filament), are not covered by the warranty.
Technical Data
Measurement principle
100 mbar
1)
… 1500 mbar
5×10
-4
… 1 mbar
1 … 100 mbar
1)
capacitance diaphragm
sensor
thermal conductance
according to Pirani
crossover range
Measurement range
(air, O
2
, CO, N
2
) 5×10
-4
… 1500 mbar
Accuracy (N
2
)
1×10
-3
… 50 mbar
50 … 950 mbar
Atmospheric pressure
(950 … 1050 mbar)
±15% of reading
±5% of reading
±2.5% of reading
Repeatability (N
2
)
±2% of reading
(1×10
-3
… 1100 mbar)
Output signal
(measurement signal)
Voltage range 0 … +9.0 V
Mesasurement range
+2.2 … +8.68 V
Voltage vs. pressure 1 V/decade, logarithmic
Output impedance
2 × 4.7 , short circuit-proof
Minimum load impedance
10 k
Response time 10 ms
Gauge identification
resistor 3 k between pin 1
and pin 5 of electrical
connector
( "Electrical Connection")
HV Adjustment
at <<10
-4
mbar
(with potentiometer <HV>)
Supply
DANGER
The gauge may only be connected to supply and
evaluation units that conform to the requirements
of a grounded protective extra-low voltage
(SELV-E according to EN 61010). The connec-
tion to the gauge has to be fused.
Supply voltage at gauge
+15 … +30 VDC
(ripple 1 V
pp
)
Power consumption
2.5 W
Fuse to be connected 1 AT
Electrical connection Hirschmann connector
type GO 6, 6 pin, male
Sensor cable 5 poles plus shielding
Cable length
150 m (5×0.25 mm
2
)
200 m (5×0.34 mm
2
)
Grounding concept
Vacuum connection to
signal ground
( "Electrical Connection")
connected via 1 M
Materials exposed to vacuum
Vacuum connection
Pirani filament
Feedthrough
Orifice
Other materials
stainless steel
tungsten
glass
stainless steel
Ni, Cu, NiFe, SnAg, glass,
Al
2
O
3
(>99.5%), AgPd
Internal volume
DN 16 ISO-KF
DN 16 CF-F
6 cm
3
8 cm
3
Admissible pressure
5 bar (absolute)
Admissible Temperatures
Storage
Operation (ambient)
Bake-out
2)
Filament temperature
–20 … +65 °C
+10 … +50 °C
80 °C
<160 °C
Relative humidity
80% at temperatures
+31 °C, decreasing to 50%
at +40 °C
Mounting orientation any
Use indoors only
altitudes up to 2000 m NN
Protection category IP 40
1)
Crossover range for air, O
2
, CO and N
2
10 mbar,
100 mbar in heavy gases.
2)
Temperature at vacuum connection with horizontal moun-
ting orientation. During bake-out, measurement range,
accuracy, and repeatability may deviate from specifi-
cations.
Vista de página 0
1 2 3 4

Resumo do Conteúdo

Página 1 - Compact Pirani Capacitance

Operating InstructionsIncl. Declaration of ConformityCompact Pirani CapacitanceGaugePCR 260BG 805 180 BE / C (2006-05)Product IdentificationIn all

Página 2 - Installation

Dimensions [mm]26.3DN 16 ISO-KFDN 16 CF-F39HV2865.31622.52043WeightDN 16 ISO-KFDN 16 CF-F 90 g120 gOutput Signal vs. Pressure1E+051E+041E+031E+021E+0

Página 3 - Deinstallation

Electrical ConnectionMake sure the vacuum connection is properly made(→ "Vacuum Connection"). If no sensor cable is available, make one acc

Página 4 - Serial number

Maintenance, RepairGauge failures due to contamination, as well asexpendable parts (filament), are not covered by thewarranty.The product requires no